Cytotoxic safety boxes Cytoculture EscoWorking with cytotoxic substances and chemotherapy drugs requires additional safety systems that are not available in standard models of Class II biological safety boxes. Since cytotoxic agents pose an extreme danger to humans and cannot be decontaminated by classical chemical decontamination methods, additional protection is required for both the operator during operation and the technician during filter changes.

Description of the cytotoxic safety room Cytoculture

ESCO Cytotoxic Safety Boxes are based on the well-known Labculture Class II Type A2 model, which meets the requirements of the new European standard for biological safety boxes EN 12469:2000.

A key feature of the CYT cytotoxic safety boxes is the presence of an additional HEPA filter installed directly under the work surface. This filter is changed safely for personnel and the environment when the box is in operation and the fan is on.

Characteristics of cytotoxic safety boxes Cytoculture

  • The conveniently located Sentinel™ Silver microprocessor-based control panel makes it easy to adjust operating parameters as needed.
  • The airflow information is continuously shown on the LCD display. Any deviations from the set parameters are accompanied by acoustic and visual
    signals.
  • Double ULPA filters and an additional secondary HEPA filter at the output.
  • The filters can be changed without exposing the environment or operating personnel to potential risks.
  • The highest level of cleanliness in the work area for personnel safety and material purity.
  • The front glass is equipped with an electric lifting system to facilitate the operator’s work.
  • The large legroom maximizes operator comfort, with a depth of 245 mm.
  • All painted surfaces have an antibacterial Isocide coating.
  • Meets the requirements of EN 12469 for biological safety boxes

A brief table comparing the models of Cytoculture cytotoxic safety boxes:

Model.

CYT-4A1 CYT-6A1
External dimensions (W x D x H) 1420 x 846 x 2190 mm 2030 x 846 x 2190 mm
Internal dimensions (W x D x H) 1270 x 620 x 670 mm 1870 x 620 x 670 mm
Height of the working slot 203 mm 203 mm
Maximum opening of the front glass 440 mm 440 mm
Air flow rate Air intake 0.45 m/s
Downward flow 0.30 m/s
Exhaust volume when using an external hood Air intake 0.45 m/s
Downward flow 0.30 m/s
Exhaust volume CBV 611m3/h 931m3/h
Constant pressure of the external hood 39 Pa 63 Pa
Typical efficiency of a ULPA filter inbound >99.9995% for a particle size of 0.3 μm according to IEST-RP-CC001.3 USA
outgoing 99.999% for particle size from 0.1 to 0.2 microns in accordance with IEST-RP-CC001.3 USA
Typical efficiency of a HEPA filter outgoing 99.999% for a particle size of 0.3 μm in accordance with IEST-RP-CC001.3 USA
Noise level 61 dB 63 dB
Fluorescent lamp intensity at zero ambient >1300 suites >1000 lux
Body material Outer surface 1.5 mm 16 gauge galvanized steel with white Isocide epoxy powder coating
Inner surface Stainless steel Type 304 with coating No. 4, thickness 1.5 mm / 16 gauge
Rated power consumption 653 W 793 W
Net weight 383 kg 500 kg
Weight during transportation 415 kg 578 kg
Overall dimensions 1560 x 930 x 2230 mm 2170 x 930 x 2230 mm